5.0  Intended use 
AMI Chest ECG (VA07) 
It is used to acquire the ECG parameters of an adult users. The acquired ECG parameters 
can be used to obtain clinical consultation from cardiologists or healthcare practitioners.  
6.0  Prerequisites 
Some basic requirements must be adopted as mentioned below to ensure better operation 
of the product. 
 
AMI Chest ECG device is designed to be operated with applications like Android, iOS 
or Windows. 
 
The Android OS version of the using computer/mobile device should be 4.0 or higher. 
 
The iOS version of the using computer/mobile device should be 7 or higher. 
 
The Windows OS version of the using computer/mobile device should be 8 or higher. 
7.0  Using Mobile Application    
   Downloading and Installing Mobile application in your computer/mobile device 
1.  Download and Install the application from Play store/Application store/Windows 
application market. 
2.  On successful installation, launch the application in the mobile device. 
 
8.0  AMI Chest ECG Measurement procedure 
1.  Position the disposable ECG electrodes on the user`s body as shown below.  
 
Note:  Kindly  replace  all  the  3  ECG  electrodes  after  24  hours  of 
usage.  While  replacing,  do  not  place  the  electrodes  on  the  same 
position. Instead, place it slightly away / adjacent to the previous 
location. 
 
2.  Ensure that the positioning spots of the human body are clean shaved.