Valeport Ltd 0804IDU User Manual

Page of 83
 
HydrINS 2 / HydrINS 2 Mini flow meter  
Installation Manual  
 
Page 28 / 83 
5.2  Method for installation at 1/8 
 
 
 
Figure 16: Diagram of installation at 1/8 diameter 
For an installation at 1/8 diameter, the distance Z
1/8
 after probe insertion 
must be equal to: 
Z
1/8 
= L - E - 1/8 DI 
 
Z
1/8
: Distance between the lower edge of the guide bar and the 
top of the pipe (mm or in) 
 
E: Pipe thickness (mm or in) 
 
D: Inside diameter of pipe (mm or in) 
 
L: Length (mm or in) – value according to model, equal to the 
distance between the lower edge of the guide bar and the two 
measurement electrodes located on the sensor 
Model 
L mm (inches) 
300 
760 (29.92) 
500 
935 (36.81) 
700 
1175 (46.26) 
1000 
1475 (58.07) 
Mini 
575 (22.64) 
Table 2: Correspondence of lengths L with the HydrINS probe model for an installation at 
1/8th diameter 
 
Z
1/8
 
1/8 DI 
DI 
Guide Bar