Zebra HW46211 User Manual

Page of 275
Page 2-34
110XiIIIPlus Maintenance Manual
13185L-002 Rev. A 1/24/06
Section 2
Operations Overview
Sensor Profile
Press the right oval key to print a graphic representation (Media Sensor Profile) of the 
changes in density between the media and the web (liner). Use the sensor profile to help 
troubleshoot media registration problems.
Refer to 
. The media sensor profile shows three conditions. The black area 
along the bottom of the profile illustrates media passing by the media sensor. When the 
level rises above the point labeled Web (black spikes), only the liner is passing by the 
sensor. When a notch or hole in the media passes by the sensor, the level rises above the 
point labeled Media. If the level remains above the Media point for longer than 
0.5 seconds, this signifies a media-out condition. The ribbon profile indicates ribbon in if 
the black level is above the point labeled Ribbon.
Figure 2-18.   Sensor Profile Sample Label
Press NEXT/SAVE to display MEDIA AND RIBBON.
Calibrate Media and Ribbon Sensor 
Press the right oval key to perform calibration.
Refer to 
 for further details.
SENSOR PROFILE
PRINT
Ribbon
Media
Web
Note • Before you begin this procedure, ensure the maximum length is set to a 
value greater than the length of the labels you are using. If the maximum length 
is set to a lower value, the calibration process assumes that continuous media 
is in the printer.
Ensure that the media type and maximum length values have been configured 
prior to performing this calibration process.
MEDIA AND RIBBON
CALIBRATE