Globtek USB charger Mains socket WR9QA1200USBNMDEURVB USB 1 x 1200 mA WR9QA1200USBNMDEURVB Data Sheet

Product codes
WR9QA1200USBNMDEURVB
Page of 12
GlobTek, Inc.
www.globtek.com
186 Veterans Drive,
Northvale, NJ 07647
Tel. (201) 784-1000
Fax (201) 784-0111
Drawing Title:  GTM41078-0605-USB, Medical Power Supply, Efficiency Level: V, Case Color/Style: GlobTek
Black, Wall Plug-in, Regulated Switchmode AC-DC, Input Rating: 100-240V~, 50-60 Hz, Interchangeable
Blades http://en.globtek.com/interchangeable-blades.php, Output Rating: 6W, 5V@1.2A, Output Configuration:
N/A mm, N/A/4 Cond, N/A, Type "A" USB “JACK”, integral to power supply enclosure, Ferrite: None, N/A, GT
Cord # C0132, Pin Out: Pin 1: +V, Pin 2-3: connected, Pin 4: DC Common, Blades/Input Cord: Q-Series
Europe, Approvals: CB, cETLus, cETLus 60601-1 3rd, China RoHS, Double Insulation, EC, GOST-R, IP20,
Level V, LPS, PSE, RoHS 2, S-Mark, Ukraine, WEEE
Model No.  GTM41078-0605-USB
Part No.  WR9QA1200USBNMDEURVB
Rev A
PROPRIETARY INFORMATION
PROPRIETARY OF GLOBTEK, INC. ANY REPRODUCTION, DISCLOSURE OR USE OF THIS DRAWING, IN WHOLE OR IN PART, IS HEREBY PROHIBITED EXCEPT AS SPECIFIED IN
WRITING BY GLOBTEK, INC.
rmünt /07/16/2014 10:12:07 AM
Page 2 of 12
POWER SUPPLY INFORMATION
TYPE:
Wall Plug-in
TECHNOLOGY:
Regulated Switchmode AC-DC
CASE COLOR:
GlobTek Black
NAMEPLATE RATED
INPUT:
100-240V~, 50-60 Hz, 0.3A
INPUT CONFIG:
Interchangeable Blades
WATTS:
6.0
VOLTS OUT:
5
CURRENT OUT (Amps):
1.2
BLADE/CORD INCLUDED: Q-EU: Q-Series Europe
BLADE/CORD INSTALLED: Q-EU: Q-Series Europe
EFFICIENCY LEVEL:
V
OUTPUT CONFIGURATION
N/A mm, N/A/4 Cond, N/A, Type "A" USB “JACK”, integral to power supply enclosure, Ferrite: None, N/A, GT Cord #
C0132
CONNECTOR PIN OUT:
Pin 1: +V, Pin 2-3: connected, Pin 4: DC Common
LABEL SPECS:
Standard GT,L-1186
PACK SPEC:
Standard with USB O/P, Individually Box
NOTES / DEVIATIONS:
DIMENSIONS ARE IN MM UNLESS SPECIFIED OTHERWISE.
ELECTRICAL SPECIFICATIONS
A) ELECTRICAL SPECIFICATIONS:
1.     Input Voltage: Specified 90-264 Vac, Nameplate rated 100-240 Vac 
2.     Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60 Hz
3. Output Power (Rated): 6Watts maximum
4. Output Regulation: +/- 5% measured at the output connector
5. Line Voltage Regulation: +/- 1% typical measured at the output connector
6. Output Ripple (Vp-p): +/-1% or 150 mV whichever is greater at nominal output voltage. Measured at 20 MHz bandwidth with 0.1 uf
ceramic capacitor in parallel with 10 uf electrolytic capacitor connected at the end of the output connector.
7. Turn-On/Turn-Off Overshoot: 5% maximum, 500 us maximum recovery time for 25% step load.
8. Turn-On Delay: 3000 mSEC maximum
9. Hold-Up Time: 8mSec maximum at nominal input and full load.
10. Inrush Current: 30A typical @ 115Vac input ; 60A typical @ 230Vac input
11. Switching Frequency: 66.5 KHz typical
B) PROTECTION
1. Over-Voltage: Protected unit will recover upon removal of fault 
2. Short Circuit: Electronically Protected, unit will recover upon removal of fault
3. Input Protection: Input line fusing
C) SAFETY
1. Dielectric Withstand Voltage: 5656Vdc from primary to secondary
2. Earth Leakage Current: N/A for Class II units, there is no PE Ground pin, so Earth Leakage current is not measured 
3.     Touch Current: Maximum allowed values: 100uA NC(Normal condition) 500uA SFC(single fault condition) 
4.     Means of  Protection: 2 x MOPP
5.     ROHS 2: Complies with EU 2011/65/EU and China SJ/T 11363-2006
D) OTHER:
1. MTBF: 200,000 hours @ 25°C ambient temperature
2. Operating Temperature: 0°C to 40°C ambient temperature
3. Humidity: 0% to 95% relative humidity
4. Storage Temperature: -40°C to 80°C