25 kHz-LCR-Meter HM8018
All data valid at 23 °C after 30 minutes warm-up.
Measurement functions
Measuring modes:
R, L, C, Θ, Q/D, |Z|
Equivalent circuits:
serial, parallel
Measuring method:
2-wire, 4-wire
Measuring ranges:
R: 0.001 Ω…99.9 MΩ 
C: 0.001 pF…99.9 mF 
L: 0.01 µH…9,999 H 
Q: 0.0001…99.9 
D: 0.0001…9.9999 
Θ: (-180.00°)…(+180.00°)
Basic accuracy:
0.2 %
Measuring frequencies:
100 Hz, 120 Hz, 1 kHz, 10 kHz, 25 kHz
Freq. Accuracy:
±100 ppm (except 120 Hz: 120.2 Hz ±100 ppm)
Measuring voltage:
0.5 V
rms
 ±10 % (unloaded)
Measuring rate:
2 measurements/second
Range changing:
automatic, manual
DC Bias voltage:
1 V ±10 %
Zero setting:
Open/short circuit compensation
Compensation limits:
Short:   R <10 Ω 
Z <15 Ω
Open:  Z >10 kΩ
Measurement accuracy
with D <0.1 or Q >10
C: A
e
 = A
f
 x A
(1 + C
x
 /C
max
 + C
min
 /C
x
L: A
e
 = A
f
 x A
(1 + L
x
 /L
max
 + L
min
 /L
x
Z: A
e
 = A
f
 (1 + Z
x
 /Z
max
 + Z
min
 /Z
x
R: A
e
 = A
f
 x A
(1 + R
x
 /R
max
 + R
min
 /R
x
A
= 1 for D <0.1
with D ≥0.1
A
d
 = ZX
1 + D
2
with the parameters
C
x
, L
x
, Z
x
, R
x
 = Measurement value 
A
f
 = 0.2 % at f = 100 Hz, 120 Hz, 1 kHz 
A
f
 = 0.3 % at f = 10 kHz 
A
f
 = 0.5 % at f = 25 kHz 
Parameter 
Auto Range
C
max
 
160 µF/f (f in kHz)
C
min
 
53 pF/f (f in kHz)
L
max
 
480 H/f (f in kHz)
L
min
 
0.16 mH/f (f in kHz)
Z
max
, R
max
 
3 MΩ
Z
min
, R
min
 
0.5 Ω
Dissipation factor accuracy:
 
Quality factor accuracy:
 
Phase angle accuracy:
Display
5-digits 7-Segment LEDs with sign
Display Parameters:
Value
% Value
Deviation
% Offset
}
 
Calculation from measurement value and 
reference value stored
Miscellaneous
The inputs are short-circuit-proof and overvoltage protected up to 100 V
dc
 
with a maximum energy consumption of 1 J.  
One configuration can be saved.
Power supply
(from mainframe):
+5 V/300 mA 
+5.2 V/50 mA 
-5.2 V/50 mA 
(∑ = 2 W)
Operating temperature:
+5…+40 °C
Storage temperature:
-20…+70 °C
Rel. humidity:
5…80 % (non condensing)
Dimensions (W x H x D)
(without 22-pole flat plug):
 
135 x 68 x 228 mm
Weight:
approx. 0.5 kg
Included in delivery: Operating manual, CD
Recommended accessories:
HZ10S  5 x silicone test lead (measurement connection in black)
HZ10R  5 x silicone test lead (measurement connection in red)
HZ10B  5 x silicone test lead (measurement connection in blue)
HZ17 
Kelvin test lead (4-wire) with probe tips
HZ18 
Kelvin test lead (4-wire) with gold plated contacts
HZ19 