Hameg HM8018 Digital-Multimeter, DMM, 2000 digits 24-8018-0200 Data Sheet
Product codes
24-8018-0200
HM8018E/100812 · C&E · Subject to change without notice · © HAMEG Instruments GmbH
®
· DQS-certified in accordance with DIN EN ISO 9001:2008, Reg.-No.: 071040 QM08
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 8000 · Fax +49 (0) 6182 800100 · www.hameg.com · info@hameg.com
25 kHz-LCR-Meter HM8018
All data valid at 23 °C after 30 minutes warm-up.
Measurement functions
Measuring modes:
R, L, C, Θ, Q/D, |Z|
Equivalent circuits:
serial, parallel
Measuring method:
2-wire, 4-wire
Measuring ranges:
R: 0.001 Ω…99.9 MΩ
C: 0.001 pF…99.9 mF
L: 0.01 µH…9,999 H
Q: 0.0001…99.9
D: 0.0001…9.9999
Θ: (-180.00°)…(+180.00°)
Θ: (-180.00°)…(+180.00°)
Basic accuracy:
0.2 %
Measuring frequencies:
100 Hz, 120 Hz, 1 kHz, 10 kHz, 25 kHz
Freq. Accuracy:
±100 ppm (except 120 Hz: 120.2 Hz ±100 ppm)
Measuring voltage:
0.5 V
rms
±10 % (unloaded)
Measuring rate:
2 measurements/second
Range changing:
automatic, manual
DC Bias voltage:
1 V ±10 %
Zero setting:
Open/short circuit compensation
Compensation limits:
Short: R <10 Ω
Z <15 Ω
Open: Z >10 kΩ
Measurement accuracy
with D <0.1 or Q >10
C: A
e
= A
f
x A
d
(1 + C
x
/C
max
+ C
min
/C
x
)
L: A
e
= A
f
x A
d
(1 + L
x
/L
max
+ L
min
/L
x
)
Z: A
e
= A
f
(1 + Z
x
/Z
max
+ Z
min
/Z
x
)
R: A
e
= A
f
x A
d
(1 + R
x
/R
max
+ R
min
/R
x
)
A
d
= 1 for D <0.1
with D ≥0.1
A
d
= ZX
1 + D
2
with the parameters
C
x
, L
x
, Z
x
, R
x
= Measurement value
A
f
= 0.2 % at f = 100 Hz, 120 Hz, 1 kHz
A
f
= 0.3 % at f = 10 kHz
A
f
= 0.5 % at f = 25 kHz
Parameter
Auto Range
C
max
160 µF/f (f in kHz)
C
min
53 pF/f (f in kHz)
L
max
480 H/f (f in kHz)
L
min
0.16 mH/f (f in kHz)
Z
max
, R
max
3 MΩ
Z
min
, R
min
0.5 Ω
Dissipation factor accuracy:
Quality factor accuracy:
Phase angle accuracy:
Display
5-digits 7-Segment LEDs with sign
Display Parameters:
Value
% Value
Deviation
% Offset
}
Calculation from measurement value and
reference value stored
Miscellaneous
The inputs are short-circuit-proof and overvoltage protected up to 100 V
dc
with a maximum energy consumption of 1 J.
One configuration can be saved.
Power supply
(from mainframe):
+5 V/300 mA
+5.2 V/50 mA
-5.2 V/50 mA
(∑ = 2 W)
Operating temperature:
+5…+40 °C
Storage temperature:
-20…+70 °C
Rel. humidity:
5…80 % (non condensing)
Dimensions (W x H x D)
(without 22-pole flat plug):
135 x 68 x 228 mm
Weight:
approx. 0.5 kg
Included in delivery: Operating manual, CD
Recommended accessories:
HZ10S 5 x silicone test lead (measurement connection in black)
HZ10R 5 x silicone test lead (measurement connection in red)
HZ10B 5 x silicone test lead (measurement connection in blue)
HZ17
HZ10S 5 x silicone test lead (measurement connection in black)
HZ10R 5 x silicone test lead (measurement connection in red)
HZ10B 5 x silicone test lead (measurement connection in blue)
HZ17
Kelvin test lead (4-wire) with probe tips
HZ18
Kelvin test lead (4-wire) with gold plated contacts
HZ19