Performance Verification
Frequency
Amplitude
300 MHz
–27 dBm (~10 mV
RMS
)
2.5 GHz
–21 dBm (~20 mV
RMS
)
3.0 GHz
–15 dBm (~40 mV
RMS
)
FCA3X20
250 MHz
–21 dBm (~20 mV
RMS
)
500 MHz
–21 dBm (~20 mV
RMS
)
18 GHz
–27 dBm (~10 mV
RMS
)
20 GHz
–21 dBm (~20 mV
RMS
)
MCA30XX
300 MHz
–33 dBm (~5 mV
RMS
)
18 GHz
–33 dBm (~5 mV
RMS
)
20 GHz
–29 dBm (~8 mV
RMS
)
27 GHz
–27 dBm (~10 mV
RMS
)
40 GHz
–23 dBm (~16 mV
RMS
)
Push the User Opt button, and then select Save Recall Recall Setup Default
to return the instrument to the factory default settings.
• This completes the Performance Verification procedure 