DS1000Z Programming Guide
:TRIGger:PATTern
Command List:
:TRIGger:PATTern:PATTern
Syntax
:TRIGger:PATTern:PATTern <pattern1>,<pattern2>,<pattern3>,<pattern4>
:TRIGger:PATTern:PATTern?
:TRIGger:PATTern:PATTern?
Description
Set or query the pattern of each channel in pattern trigger.
Parameter
Name
Type
Range
Default
<pattern1>
Discrete
{H|L|X|R|F}
X
<pattern2>
Discrete
{H|L|X|R|F}
X
<pattern3>
Discrete
{H|L|X|R|F}
X
<pattern4>
Discrete
{H|L|X|R|F}
X
Explanation
H: set the pattern of the channel selected to “H”, namely the voltage level is higher
than the threshold level of the channel.
L: set the pattern of the channel selected to “L”, namely the voltage level is lower
than the threshold level of the channel.
X: set the pattern of the channel selected to “Don’t Care”, namely this channel is not
used as a part of the pattern. When all the channels in the pattern are set to “Don’t
Care”, the oscilloscope will not trigger.
R/F: set the pattern to the rising or falling edge of the channel selected.
You can not to specify patterns of two channels to rising edge or falling edge at the
same time. If one edge item is currently defined and then another edge item is
defined in another channel in the pattern, the former edge item defined will be
replaced by X.
The four parameters from left to right in the command set the pattern of CH1, CH2,
CH3 and CH4 respectively.
Return
Format
The query returns the patterns of all channels.
Example
:TRIGger:PATTern:PATTern H,R,L,X