  DS1000Z Programming Guide 
:TRIGger:TIMeout:TIMe 
Syntax 
:TRIGger:TIMeout:TIMe <NR3> 
:TRIGger:TIMeout:TIMe? 
Description 
Set or query the timeout time in timeout trigger. The default unit is s.   
Parameter 
Name 
Type 
Range 
Default 
<NR3> 
Real 
16ns to 10s 
16ns 
 
Return 
Format 
The query returns the timeout time in scientific notation.   
Example 
:TRIGger:TIMeout:TIMe 0.002   /*Set the timeout time to 2ms*/ 
:TRIGger:TIMeout:TIMe?   /*The query returns 2.000000e-03*/ 
Related 