Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S User Manual
Product codes
MSO1074Z-S
RIGOL
2-44
DS1000Z Programming Guide
:MATH:OPTion:THReshold1
Syntax
:MATH:OPTion:THReshold1 <thre>
:MATH:OPTion:THReshold1?
:MATH:OPTion:THReshold1?
Description
Set or query the threshold level of the source A in logic operation. The default unit is V.
Parameter
Name
Type
Range
Default
<thre>
Real
(-4*Vertical Scale – Vertical Offset)~
(4*Vertical Scale – Vertical Offset)
0.00V
Explanation
This command is only applicable to the logic operations (A&&B, A||B, A^B and !A).
Vertical Scale refers to the vertical scale of source A
Vetical Offset refers to the vertical offset of source A
The step is Vertical Scale/25
Return
Format
The query returns the threshold level in scientific notation.
Example
:MATH:OPTion:THReshold1 0.8 /* Set the threshold level to 800mV*/
:MATH:OPTion:THReshold1? /*The query returns 8.000000e-01*/
Related
Command
:MATH:OPTion:THReshold2
Syntax
:MATH:OPTion:THReshold2 <thre>
:MATH:OPTion:THReshold2?
:MATH:OPTion:THReshold2?
Description
Set or query the threshold level of the source B in logic operation. The default unit is V.
Parameter
Name
Type
Range
Default
<thre>
Real
(-4*Vertical Scale –Vertical Offset)~
(4*Vertical Scale – Vertical Offset)
0.00V
Explanation
This command is only applicable to the logic operations (A&&B, A||B, A^B and !A).
Vertical Scale refers to the vertical scale of source B
Vetical Offset refers to the vertical offset of source B
The step is Vertical Scale/25
Return
Format
The query returns the threshold level in scientific notation.
Example
:MATH:OPTion:THReshold2 0.5
Related
Command