Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S User Manual

Product codes
MSO1074Z-S
Page of 264
 
RIGOL 
DS1000Z User’s Guide
 
 
4-7  
 
Memory Depth 
 
Memory depth refers to the number of waveform points that the oscilloscope can 
store in a single trigger sample and it reflects the storage ability of the sample 
memory. DS1000Z provides up to 24 Mpts memory depth (option) and 12 Mpts 
standard memory depth.   
T
Trigger Point
Pre-sample
Delayed Sample
Memory Depth
 
 
The relation of memory depth, sample rate and waveform length fulfills the equation 
below:   
 
Memory depth = sample rate (Sa/s) × waveform length (s/div × div) 
 
Thus, in the same time base, the higher memory depth can ensure a higher sample 
rate. 
 
Press Acquire  Mem Depth, use 
  to switch to the desired memory depth (the 
default is auto) and then press down the knob to select the option. You can also 
press Mem Depth continuously to switch the memory depth. 
 
When a single channel is enabled, the memory depths available include Auto, 
12kPoints, 120kPoints, 1200kPoints, 12MPoints. In “Auto” mode, the oscilloscope 
selects the memory depth automatically according to the current sample rate.   
 
When two channels are enabled, the memory depths available include Auto, 6kPoints, 
60kPoints, 600kPoints, 6MPoints. In “Auto” mode, the oscilloscope selects the 
memory depth automatically according to the current sample rate.