Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S User Manual

Product codes
MSO1074Z-S
Page of 264
 
RIGOL 
DS1000Z User’s Guide
 
 
5-5  
 
Trigger Coupling 
 
Trigger coupling decides which kind of components will be transmitted to the trigger 
circuit. Please distinguish it with 
.   
  DC: allow DC and AC components into the trigger path.  
  AC: block all the DC components and attenuate signals lower than 75 kHz. 
  LF Reject: block the DC components and reject the low frequency components 
(lower than 75 kHz). 
  HF Reject: reject the high frequency components (higher than 75 kHz). 
 
Press MENU  Setting  Coupling in the trigger control area (TRIGGER) at the 
front panel to select the desired coupling type (the default is DC). Note that trigger 
coupling is only valid in edge trigger.