Rigol DS2072A 2-channel oscilloscope, Digital Storage oscilloscope, DS2072A User Manual
Product codes
DS2072A
RIGOL
DS2000A User’s Guide
6-20
4. Phase Measurement Setting
Specify the source A and source B in the measurement items “Phase A→B ”
and “Phase A→
and “Phase A→
B ”.
Press Measure Meas.Setting Type “Phase” and then press SourceA
and SourceB to set the two channel sources (CH1 or CH2) of phase
measurement respectively.
and SourceB to set the two channel sources (CH1 or CH2) of phase
measurement respectively.
5. Threshold Measurement Setting
Specify the vertical level (in percentage) being measured in the analog channel.
Measurements of all the time, delay and phase parameters will be influenced by
this setting.
Press Measure Meas.Setting Type “Threshold” and then:
Press Max and use
Measurements of all the time, delay and phase parameters will be influenced by
this setting.
Press Measure Meas.Setting Type “Threshold” and then:
Press Max and use
to set the maximum value of the measurement.
Reducing the maximum value to the current “Mid” will automatically reduce
the “Mid” to keep it lower than the maximum value. The default is 90% and
the range available is from 7% to 95%.
the “Mid” to keep it lower than the maximum value. The default is 90% and
the range available is from 7% to 95%.
Press Mid and use
to set the middle value of measurement. The
middle value is limited by the settings of “Max” and “Min”. The default is 50%
and the range available is from 6% to 94%.
and the range available is from 6% to 94%.
Press Min and use
to set the minimum value of the measurement.
Increasing the minimum value to the current “Mid” will automatically
increase the “Mid” to keep it higher than the minimum value. The default is
10% and the range available is from 5% to 93%.
increase the “Mid” to keep it higher than the minimum value. The default is
10% and the range available is from 5% to 93%.