Rigol DS2072A 2-channel oscilloscope, Digital Storage oscilloscope, DS2072A User Manual

Product codes
DS2072A
Page of 284
 
RIGOL 
DS2000A User’s Guide
 
 
12-11  
 
7.  Align Phase 
The align phase operation is for the two channels. Pressing AlignPhase will 
re-configure the two channels to output according to the preset frequency and 
phase. For two signals of which the frequencies are the same or are multiples, 
this operation can align their phase. Acquire the waveforms of the two channels 
using the oscilloscope and display the waveforms stably. After switching the 
channel status, the phase deviation between the two waveforms is changed. At 
this point, press AlignPhase and the phase deviation shown on the oscilloscope 
will restore to the current phase deviation between the two waveforms 
automatically.   
 
8.  Modulation 
Press Modulation to turn on or off the modulation function. For detailed 
introduction of the modulation function, please refer to 
. 
 
9.  Set the impedance 
Press Impedance to set the output impedance of the signal generator. It can 
be set to “HighZ” or “50Ω”. 
 
 
To Output Arbitrary Waveform 
 
DS2000A allows users to define their own arbitrary waveform and save it in the 
internal or external memory. At most 10 arbitrary waveforms can be stored in the 
internal memory. The user-defined waveform can contain 1 to 16384 data points 
(namely 1 pts to 16 kpts). 
 
1.  Source 
Press Source to turn on or off the signal output. 
 
2.  Set the frequency 
Press Frequency to set the frequency of the signal. The range is from 100 mHz 
to 25 MHz.   
Method 1: 
Rotate 
  or the navigation knob to set the desired frequency value.