Test and Ouput 
 
Before the test, you can use the following method to set the output mode of the test 
results.   
Press Msg Display to select “ON” or “OFF”. When “ON” is selected, the test results 
will be displayed at the upper right corner of the screen.   
 
 
Press Stop On Outp to select “ON” or “OFF”.   
  ON: when failed waveforms are detected, the oscilloscope will stop the test and 
enter the “STOP” state. At this point, the results of the test remain the same on 
the screen (if display is turned on) and only one negative pulse is output from 
the [Trigger Out] connector (if enabled) at the rear panel. 
  OFF: the oscilloscope will continue with the test even though failed waveforms 
are detected. The test results on the screen will update continuously and a pulse 
will be output from the [Trigger Out] connector at the rear panel each time a 
failed waveform is detected.