Rigol DS2302A-S 2-channel oscilloscope, Digital Storage oscilloscope, DS2302A-S User Manual

Product codes
DS2302A-S
Page of 284
RIGOL 
  DS2000A User’s Guide 
5-40 
USB Trigger (Option) 
 
Trigger on the SOP, EOP, RC, Suspend and Exit Suspend of the data packet on the 
differential USB data cable (D+ and D-). This trigger supports USB Low Speed and 
Full Speed. The figure below shows the USB data transmission protocol.   
 
 
 
Trigger Type: 
Press Type to select “USB”. At this point, the trigger setting information as shown in 
the figure below is displayed at the upper right corner of the screen. 
 
 
Source Selection: 
Press D+ and D- to specify data sources for D+ and D- data cables respectively. 
They can be set to CH1 or CH2 and the current trigger sources are displayed at the 
upper right corner of the screen.   
 
Signal Speed: 
Press Speed to select “Low Speed” (1.5 Mb/s) or “Full Speed” (12 Mb/s).   
 
Trigger Condition: 
Press When to select the desired trigger condition. 
  SOP: trigger at the sync bit
 
at the start of the data packet (SOP). 
  EOP: trigger at the end of the SEO portion of the EOP of the data packet.   
  RC: trigger when SEO is greater than 10 ms.   
  Suspend: trigger when the idle time of the bus is greater than 3 ms.   
  ExitSuspend: trigger when the bus exits from idle state for more than 10 ms. 
 
Trigger Mode: 
Press Sweep to select the 
 (page 
) under this trigger type as auto,