RIGOL 
DS2000A User’s Guide
 
 
6-17  
 
Other Parameters 
 
1. 
 
Area: the area of the whole waveform within the screen and the unit is 
voltage-second. The area meadured above the zero reference (namely the 
vertical offset) is positive and the area measured below the zero reference is 
negative. The area measured is the algebraic sum of the area of the whole 
waveform within the screen.   
 
2. 
 
Period Area: the area of the first period of waveform on the screen and 
the unit is voltage-second. The area above the zero reference (namely the 
vertical offset) is positive and the area below the zero reference is negative. The 
area measured is the algeraic sum of the area of the whole period waveform. 
Note that when the waveform on the screen is less than a period, the period 
area measured is 0.