  DS2000A User’s Guide 
12-2 
To Output Basic Waveform 
To Output Sine 
 
Press Src1 Setup to open the waveform setting menu. Press Wave and select 
“Sine”. At this point, you can set the sine parameters.   
 
1.  Source 
Press Source to turn on or off the signal output. 
 
2.  Set the frequency 
Press Frequency to set the frequency of the signal. The range is from 100 mHz 
to 25 MHz. Note that for different waveforms, the frequency ranges are different. 
The frequency range of square and pulse is from 100 mHz to 15 MHz; the 
frequency range of ramp is from 100 mHz to 100 kHz. 
Method 1: 
Rotate 
  or the navigation knob to set the desired frequency value.   
Method 2: 
Press down 
  and the numeric keyboard is displayed. Use the knob to input 
the desired frequency value and unit. 
 
3.  Set the amplitude 
Press Amplitude to set the amplitude of the signal. When the impedance is set 
to HighZ, the range is from 20 mVpp to 5 Vpp; when the impedance is set to 50 
Ω, the range is from 10 mVpp to 2.5 Vpp. For the setting method, please refer to 
the introduction i
. 
 
4.  Set the DC offset voltage 
Press Offset to set the DC offset of the signal. When the impedance is HighZ, 
the range is from (-2.5 V+the current amplitude/2) to (2.5 V-the current 
amplitude/2); when the impedance is set to 50 Ω, the range is from (-1.25 V+the 
current amplitude/2) to (1.25 V-the current amplitude/2). 
Method 1: 
Rotate 
  or the navigation knob to set the desired DC offset voltage. 
Method 2: 
Press down 
  and the numeric keyboard is displayed. Use the knob to input