  DS2000A User’s Guide 
12-12 
Method 2: 
Press down 
  and the numeric keyboard is displayed. Use the knob to input 
the desired frequency value and unit. 
 
3.  Set the amplitude 
Press Amplitude to set the amplitude of the signal. When the impedance is set 
to HighZ, the range is from 20 mVpp to 5 Vpp; when the impedance is set to 50 
Ω, the range is from 10 mVpp to 2.5 Vpp. For the setting method, please refer to 
the introduction i
.   
 
4.  Set the DC offset voltage 
Press Offset to set the DC offset of the signal. When the impedance is HighZ, 
the range is from (-2.5 V+the current amplitude/2) to (2.5 V-the current 
amplitude/2); when the impedance is set to 50 Ω, the range is from (-1.25 V+the 
current amplitude/2) to (1.25 V-the current amplitude/2). 
Method 1: 
Rotate 
  or the navigation knob to set the desired DC offset voltage. 
Method 2: 
Press down 
  and the numeric keyboard is displayed. Use the knob to input 
the desired offset value and unit.   
Note: The offset value contains three effective digits. For example, when the 
amplitude is set to 10 mVpp (the impedance is 50 Ω), the offset range calculated 
is from -1.245 V to 1.245 V, while the actual offset range is -1.24 V to 1.24 V. 
 
5.  Set the start phase 
Press StartPhase to set the start phase of the signal. The range is from 0°to 
360°. 
Method 1: 
Rotate 
  to set the desired start phase. 
Method 2: 
Press down 
  and the numeric keyboard is displayed. Use the knob to input 
the desired value and unit. 
 
6.  Align Phase 
The align phase operation is for the two channels. Pressing AlignPhase will 
re-configure the two channels to output according to the preset frequency and 
phase. For two signals of which the frequencies are the same or are multiples,