DS2000A User’s Guide
 
 
7-5  
 
RS232 Decoding (Option) 
 
RS232 serial bus consists of the transmitting data line (TX) and the receiving data 
line (RX).   
 
DeviceA
DeviceB
Tx
Rx
Rx
Tx
 
 
The industry standard of RS232 uses “Negative Logic”, namely high level is logic “0” 
and low level is logic “1”.   
1
0
0
0
0
1 1
1
1
1
 
 
By default, RS232 uses LSB (Least Significant Bit) transmission sequence, namely the 
lowest bit of the data is transmitted first. While for MSB (Most Significant Bit), the 
highest bit of the data is transmitted first.   
Bit
0
Bit
5
Bit
2
Bit
3
Bit
6
Bit
1
Bit
7
Bit
4
t
        
Bit
7
Bit
2
Bit
5
Bit
4
Bit
1
Bit
6
Bit
0
Bit
3
t
 
       Endian (LSB)                          Endian (MSB) 
 
In RS232, baud rate is used to represent the transmitting rate (namely bits per 
second) of the data. The commonly used baud rates include 2400 bps, 4800 bps, 
9600 bps, 19200 bps, 38400 bps, 56000 bps and 115200 bps. 
 
In RS232, you need to set the start bit, data bits, check bit (optional) and stop bit of 
each frame of data.