  DS2000A User’s Guide 
9-2 
To Enable Pass/Fail Test 
 
Press Utility  Pass/Fail  Enable Test to select “ON”. Note that when the time 
base is in X-Y mode, the Pass/Fail test function can not be enabled.   
 
To start testing, press Enable Test and select “ON”. Then, press Operate to select 
“ ” to start testing and select “■” to stop testing.   
 
You can select the signal source, set the test mask range, create mask as well as 
save and load the test mask. For details, please refer to the following introductions.   
 
 
To Select Source 
 
Press Source to select the channel (CH1 or CH2) to be tested and only channels 
enabled can be selected. During the test, the oscilloscope will judge whether each 
frame of waveform in the source complies with the current test mask and those 
waveforms pass through the mask area (blue area) is considered as failed.   
 
 
Mask Range 
 
Users can define their desired test masks. 
 
Press MaskRange  MaskRange to select “Screen Region” or “Cursor Region” for 
test. When “Cursor Region” is selected, two gray cursor lines appear on the screen. 
At this point, press CursorA and CursorB; then, use 
  to adjust the positions of 
the two cursor lines respectively. Or, press CursorAB and use 
  to adjust the 
positions of cursor A and curcor B at the same time. Note that you can press down 
  continuously to switch the current cursor. 
 
Press X Mask and Y Mask, rotate 
  and the mask lines appear on the screen as 
shown in the figure below. Press Create Mask to apply the mask currently created. 
The horizontal and vertical adjustment ranges are 0.02 div to 4.0 div and 0.04 div to