Rigol DS2102A 2-channel oscilloscope, Digital Storage oscilloscope, DS2102A User Manual

Product codes
DS2102A
Page of 284
 
RIGOL 
DS2000A User’s Guide
 
 
3-5  
 
According to sinθ=A/B or C/D (wherein, θ is the phase deviation angle between 
the two channels and the definitions of A, B, C and D are as shown in the figure 
above), the phase deviation angle is obtained, that is: 
θ=±arcsin (A/B) or ±arcsin( C/D) 
If the principal axis of the ellipse is within quadrant I and III, the phase deviation 
angle obtained should be within quadrant I and IV, namely within (0 to π/2) or (3π/2 
to 2π). If the principal axis of the ellipse is within quadrant II and IV, the phase 
deviation angle obtained should be within quadrant II and III, namely within (π/2 to 
π) or (π to 3π/2).   
 
X-Y function can be used to measure the phase deviation occurred when the signal 
under test passes through a circuit network. Connect the oscilloscope to the circuit to 
monitor the input and output signals of the circuit.   
 
Application example: measure the phase deviation of the input signals of two 
channels. 
 
Method 1: Use Lissajous method 
 
1.  Connect a sine signal to CH1 and then connect a sine signal with the same 
frequency and amplitude but a 90° phase deviation to CH2.   
2.  Press AUTO and enable X-Y mode. Rotate Horizontal 
  SCALE to ajdust the 
sample rate properly to get better Lissajous figure for better observation and 
measurement.   
3.  Roate VERTICAL    POSITION of CH1 and CH2 to display the signals at the 
center of the screen and roate VERTICAL    SCALE of CH1 and CH2 to make 
the signals easy to observe. At this point, the circle as shown in the figure below 