RIGOL 
DS2000A User’s Guide
 
 
12-3  
 
the desired offset value and unit.   
Note: The offset value contains three effective digits. For example, when the 
amplitude is set to 10 mVpp (the impedance is 50 Ω), the offset range calculated 
is from -1.245 V to 1.245 V, while the actual offset range is -1.24 V to 1.24 V. 
 
5.  Set the start phase 
Press StartPhase to set the start phase of the signal. The range is from 0°to 
360°. 
Method 1: 
Rotate 
  to set the desired start phase. 
Method 2: 
Press down 
  and the numeric keyboard is displayed. Use the knob to input 
the desired value and unit.   
 
6.  Align Phase 
Pressing AlignPhase will re-configure the two channels to output according to 
the preset frequency and phase. For two signals of which the frequencies are the 
same or are multiples, this operation can align their phase. Acquire the 
waveforms of the two channels using the oscilloscope and display the waveforms 
stably. After switching the channel status, the phase deviation between the two 
waveforms is changed. At this point, press AlignPhase and the phase deviation 
shown on the oscilloscope will restore to the current phase deviation between 
the two waveforms automatically.   
 
7.  Modulation 
Press Modulation to turn on or off the modulation function. For detailed 
introduction of the modulation function, please refer to 
. 
 
8.  Set the impedance 
Press Impedance to set the output impedance of the signal generator. It can be 
set to “HighZ” or “50Ω”.