RIGOL 
  DS2000A User’s Guide 
12-18 
 
After selecting AM, you can set the AM frequency and AM depth of the 
modulating waveform.   
 
AM Frequency: press AM_Freq to set the frequency of the modulating 
waveform. The range is from 1 Hz to 50kHz. For the setting method, please 
refer to the introduction in 
.   
 
AM Depth: the modulation depth refers to the strength of the AM and is 
expressed in percentage. Press AM_Depth to set the modulation depth of 
the modulating waveform. The range is from 0% to 120%. When it is set to 
0%, the output amplitude is half of the carrier amplitude. When it is set to 
100%, the output amplitude equals the carrier amplitude. When it is set to 
a value greater than 100%, envelop distortion will occur which must be 
avoided in actual circuit; at this point, the output of the instrument will not 
exceed 5 Vpp (the load is 50 Ω). 
 
  FM: the frequency of the carrier waveform changes with that of the 
modulating waveform.   
Modulating Waveform 
Carrier Waveform 
Modulated Waveform