Rigol DS2202A-S 2-channel oscilloscope, Digital Storage oscilloscope, DS2202A-S User Manual

Product codes
DS2202A-S
Page of 284
 
RIGOL 
DS2000A User’s Guide
 
 
1-7  
 
Tip 
 
The signal output from the probe compensation connector can only be used for 
probe compensation adjustment and can not be used for calibration. 
 
 
Probe Compensation 
 
When the probes are used for the first time, you should compensate the probes to 
match the input channels of the oscilloscope. Non-compensated or poorly 
compensated probes may cause measurement inaccuracy or error. The probe 
compensation procedures are as follows. 
 
1.  Perform steps 1, 2, 3 and 4 o
 in the previous section.   
2.  Check the waveforms displayed and compare them with the following. 
         
 
 
 
 
Figure 1-8 Probe Compensation 
 
3.  Use a nonmetallic driver to adjust the low-frequency compensation adjustment 
hole on the probe until the waveform displayed is as the “Perfectly compensated” 
in the figure above.   
 
 
 
 
 
Over compensated        Perfectly compensated     Under compensated