  DS2000A User’s Guide 
2-8 
Delay Calibration   
 
When using an oscilloscope for actual measurement, the transmission delay of the 
probe cable may bring relatively greater error (zero offset). DS2000A allows users to 
set a delay time to calibrate the zero offset of the corresponding channel. Zero offset 
is defined as the offset of the crossing point of the waveform and trigger level line 
relative to the trigger position, as shown in the figure below.   
 
 
 
Press CH1 Delay-Cal and use 
  to set the desired delay time. The range 
available is from -200 ns to 200 ns. 
Note: This parameter is related to the instrument model and the current horizontal 
scale setting. The larger the horizontal scaleis, the larger the scale will be.