 X-Stream DSO interface capabilities include the following IEEE 488.1 definitions: 
AH1  Complete Acceptor Handshake 
DC1 
Complete Device Clear Function 
SH1 
Complete Source Handshake 
DT1 
Complete Device Trigger 
L4 
Partial Listener Function 
PP1 
Parallel Polling: remote configurable 
T5 
Complete Talker Function 
C0 
No Controller Functions 
SR1 
Complete Service Request Function 
E2 
Tri-state Drivers 
RL1 
Complete Remote/Local Function 
 
 
ADDRESS 
Every device on the GPIB has an address. To address the X-Stream DSO, set the remote control port to GPIB 
by means of  the scope’s front panel UTILITIES button and on-screen menus.  
If  you address the X-Stream DSO to talk, it will remain in that state until it receives a universal untalk 
command (UNT), its own listen address (MLA), or another oscilloscope’s talk address. 
If  you address the X-Stream DSO to listen, it will remain configured to listen until a universal unlisten 
command (UNL), or its own talker address (MTA), is received. 
To avoid conflicts, use the general Unlisten and Untalk commands before setting up the talker and listener 
states. 
The following characters are used in GPIB to control talking and listening: 
ASCII 63 = ? 
  General Unlisten    
ASCII 95 = _ 
 General Untalk 
ASCII 32 = Space 
Base Listen Address  
ASCII 64 = @ 
 Base  Talk Address