Lecroy WaveSurfer 3054 4-channel oscilloscope, Digital Storage oscilloscope, WaveSurfer 3054 User Manual

Product codes
WaveSurfer 3054
Page of 305
P
A R T  
O
N E
:
 
A B O U T
 
R E M O T E
 
C O N T R O L        
58
  
ISSUED: February 2005
 
WM-RCM-E Rev D
 
CALCULATE A DATA POINT’S HORIZONTAL POSITION 
Each vertical data value has a corresponding horizontal position, usually measured in time or frequency units. 
The calculation of  this position depends on the type of  waveform. Each data value has a position, i, in the 
original waveform, with i = 0 corresponding to the first data point acquired. The descriptor parameter 
HORUNIT gives a string with the name of  the horizontal unit. 
Single Sweep waveforms: x[i] = HORIZ_INTERVAL x i + HORIZ_OFFSET. For acquisition waveforms 
this time is from the trigger to the data point in question. It will be different from acquisition to acquisition 
since the HORIZ_OFFSET is measured for each trigger. In the case of  the data shown above this means: 
HORIZ_INTERVAL =  1e-08 from the floating point number 322b cc77 at Byte 194 
HORIZ_OFFSET =  -5.149e-08 from the double precision floating point number be6b a4bb 51a0 69bb 
at Byte 198 
HORUNIT = S =  seconds from the string 5300 ... at Byte 262 
This gives:  
x[0] =  -5.149e-08 S  
x[1] =  -4.149e-08 S.