Rigol DS1204B 4-channel oscilloscope, Digital Storage oscilloscope, DS1204B Data Sheet

Product codes
DS1204B
Page of 7
 
Specifications 
All specifications apply to the DS1000B Series Oscilloscopes unless noted otherwise. To meet these 
specifications, two conditions must first be met: 
 
  The instrument must have been operating continuously for thirty minutes within the specified 
operating temperature. 
  Must perform Self Calibration operation, accessible through the Utility menu, if the operating 
temperature changes by more than 5°C. 
 
All specifications are guaranteed unless noted “typical”. 
 
 
Technical Specifications 
 
Acquisition 
Sample Modes 
Real-Time Sample 
Equivalent Sample 
Sample Rate 
2 GSa/s (half channel 
[1]
1 GSa/s (each channel) 
50 GSa/s
[2]
 
Averages 
A waveform will be displayed one time while all the channels finish N times 
sample, N could be selectable from 2, 4, 8, 16, 32, 64, 128 and 256 
Inputs 
Input Coupling 
DC, AC, GND 
Input Impedance 
1MΩ± 2.0%   
The input capacity is 18pF± 3pF 
Probe Attenuation 
Factors 
0.001X, 0.01X, 0.1X, 1X, 2X, 5X, 10X, 20X, 50X, 100X, 200X, 500X, 1000X 
Maximum Input 
Voltage 
Maximum Input Voltage of the analog channel: 
CAT I 300Vrms, 1000Vpk; transient overvoltage 1000Vpk   
CAT II 100Vrms, 1000Vpk   
RP2200 10:1, CAT II 300Vrms   
RP3200 10:1, CAT II 300Vrms   
RP3300 10:1, CAT II 300Vrms 
Time Delay between 
Channel (typical) 
500ps 
Horizontal 
Sample Rate Range 
3.65Sa/s-2GSa/s (Real-Time), 3.65Sa/s-50GSa/s (Equivalent-time) 
Waveform 
Interpolation 
Sin(x)/x 
Record Length 
Up to 16k samples for half channel
[1]
 
8k samples for each channel   
Scanning Speed 
Range 
(Sec/div) 
1ns/div~50s/div, DS1204B   
2ns/div~50s/div, DS1104B 
5ns/div~50s/div, DS1074B 
1-2-5 Sequence   
Sample Rate and   
Delay Time Accuracy 
± 50ppm (any time interval ≥1ms) 
Delta Time 
Measurement 
Accuracy   
(Full Bandwidth) 
Single: ± (1 sample interval + 50ppm ×  reading + 0.6 ns)   
>16 averages: ± (1sample interval + 50ppm ×  reading + 0.4 ns) 
Vertical 
A/D Converter 
8-bit resolution, all channels sample simultaneously 
Volts/div Range 
2mV/div-10V/div at input BNC 
Offset Range   
± 40V(245mV/div~10V/div)