Pulse Width Trigger 
Pulse trigger occurs according to the width of pulse. The abnormal signals can be 
detected through setting up the pulse width condition.   
Figure 2-41    Table 2-22 The Trigger menu (Page 1/2)   
Menu 
Settings 
Comments 
Source 
CH1 
CH2 
CH3 
CH4 
EXT 
 
EXT/5 
Select CH1 as trigger source. 
Select CH2 as trigger source. 
Select CH3 as trigger source 
Select CH4 as trigger source l 
Select  EXT  TRIG  as  trigger 
source 
Select attenuated EXT TRIG/5 
as trigger source. 
When 
(+Pulse width more than) 
  (+Pulse width less than) 
(+Pulse width equal to) 
  (-Pulse width less than) 
(-Pulse width more than) 
(-Pulse width equal to) 
To select pulse condition. 
Settings 
 
  <Width>
 
To set required pulse width. 
Figure 2-42    Table 2-23 The Trigger menu (Page 2/2)   
Menu 
Settings 
Comments 
Sweep 
Auto 
Normal 
Single 
 
Acquire waveform even no trigger occurred. 
Acquire waveform when trigger occurred. 
When  trigger  occurs,  acquire  one  waveform 
and then stop. 
Set Up     
To go to Set Up menu, see table 2-36. 
 
 
 
Note:
 The Pulse width adjust range is 20ns ~ 10s. When the condition is met, it will 
trigger and acquire the waveform.