Rigol DS1104Z-S 4-channel oscilloscope, Digital Storage oscilloscope, DS1104Z-S User Manual

Product codes
DS1104Z-S
Page of 264
 
RIGOL 
DS1000Z User’s Guide
 
 
10-3  
 
Tip 
 
During waveform playback, RUN/STOP can be used to switch between 
playback and pause. Each time SINGLE is pressed, the Current Frame 
moves one frame forward.   
 
 
Record Setting 
 
You can press Record Opt to set the following parameters before waveform record. 
 
1.  Interval 
Press Interval to set the time interval between frames in waveform recording 
and the range available is from 100 ns to 10 s。 
 
2.  Record Length 
Press Length to set the number of frame currently recorded and the range 
available is from 1 to the maximum number of frames that can be recorded 
currently. Press Set Max to set the number of frame to the maximum number of 
frames that can be recorded. 
 
3.  Maximum Number of Frames 
The menu shows the maximum number of frames that can be recorded 
currently. 
As the capacity of the waveform memory is fixed, the more the number of points 
each frame of waveform has, the less the number of waveform frames can be 
recorded. Thus, the maximum end frame of waveform record is decided by the 
“Memory Depth” currently selected. Please refer to the instruction in Memory 
Depth
 to select the desired memory depth. 
 
4.  Beep 
 
: you can not hear the sound of the beeper when the waveform recording is 
over. 
 