DS1000Z User’s Guide 
5-2 
Trigger Source 
 
Press MENU  Source in the trigger control area (TRIGGER) at the front panel to 
select the desired trigger source. Siganls input from CH1-CH4 and AC Line can all be 
used as trigger source.   
 
 
Analog channel input: 
Signals input from analog channels CH1-CH4 can all be used as the trigger source. 
No matter whether the input of the channel selected is enabled, the channel can 
work normally. 
 
AC line: 
The trigger signal is obtained from the AC power input of the oscilloscope. This kind 
of signals can be used to display the relationship between signal (such as illuminating 
device) and power (power supply device). For example, to stably trigger the 
waveform output from the transformer of a transformer substation, which is mainly 
used in related measurement of the power industry.