Rigol DS1104Z-S 4-channel oscilloscope, Digital Storage oscilloscope, DS1104Z-S User Manual
Product codes
DS1104Z-S
RIGOL
DS1000Z User’s Guide
5-20
Trigger Condition:
Press When to select the desired trigger condition.
>: trigger when the duration of the pattern is greater than the preset time.
>: trigger when the duration of the pattern is greater than the preset time.
Press Time to set the duration of duration trigger and the range is from 8 ns to
10 s.
10 s.
<: trigger when the duration of the pattern is lower than the preset time. Press
Time to set the duration of duration trigger and the range is from 8 ns to 10 s.
<>: trigger when the duration of the pattern is lower than the upper limit of the
preset time and greater than the lower limit of the preset time. Press Upper
Limit to set the upper limit of the duration of duration trigger and the range is
from 16 ns to 10 s. Press Lower Limit to set the lower limit of the duration of
duration trigger and the range is from 8 ns to 9.99 s. Note that the time lower
limit must be lower than the time upper limit.
Trigger Mode:
) under this trigger type as auto,
normal or single. The corresponding status light of the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.
Trigger Level:
Use TRIGGER LEVEL to modify the level. For details, please refer to the
description of “
Trigger Level:
Use TRIGGER LEVEL to modify the level. For details, please refer to the
description of “
” on page
. Note that the trigger level of each
channel needs to be set respectively.