DESCRIPTION
This command sets the waveform type to a Modulated waveform. The command also can set or retrieve
modulated waveform parameters.
modulated waveform parameters.
COMMAND
SYNTAX
<channel>: MoDulateWaVe <parameter>
<channel>:= {C1,C2}
<parameter>:= {A parameter from the following table.}
Parameters
Value
Description
STATE
<state>
Sets the waveform type to a Modulated waveform or sets specific Modulated waveform
parameters.
parameters.
AM, SRC
<src>
Sets the Modulated waveform to an AM type.
AM, MDSP
<mod wave shape>
Sets the Modulated waveform to an AM type. Only applicable when the signal source is set to
INT.
INT.
AM, FRQ
<am frequency>
Sets the frequency of an AM type waveform. Only applicable when the signal source is set to
INT.
INT.
AM, DEPTH
<depth>
Sets the depth of an AM type waveform. Only applicable when the signal source is set to INT.
DSBAM, SRC
<src>
Sets the signal source of a DSBAM waveform.
DSBAM,
MDSP
MDSP
<mod wave shape>
Sets the wave shape of the DSBAM waveform. Only applicable when the signal source is set to
INT.
INT.
DSBAM, FRQ
<dsbam frequency>
Sets the frequency of a DSBAM waveform. Only applicable when the signal source is set to
INT.
INT.
FM, SRC
<src>
Sets the signal source of an FM waveform.
FM, MDSP
<mod wave shape>
Sets the wave shape of a FM waveform. Only applicable when the signal source is set to INT.
FM, FRQ
<fm frequency>
Sets the frequency of an FM waveform. Only applicable when the signal source is set to INT.
FM, DEVI
<fm frequency offset>
Sets the frequency offset of an FM waveform. Only applicable when the signal source is set to
INT.
INT.
PM, SRC
<src>
Sets the signal source of a PM waveform.
PM, MDSP
<mod wave shape>
Sets the wave shape of a PM waveform. Only applicable when the signal source is set to INT.
PM, FRQ
<pm frequency>
Sets the frequency of a PM waveform. Only applicable when the signal source is set to INT.
PWM, FRQ
<pwm frequency>
Sets the frequency of a PWM waveform. Only applicable when the signal source is set to INT.
PWM, DEVI
<pwm devi>
Sets the duty cycle deviation. Only applicable when the carrier is a Pulse waveform.
PWM, MDSP
<mod wave shape>
Sets the wave shape of a PWM waveform. Only applicable when the carrier is a Pulse
waveform.
waveform.
PWM, SRC
<src>
Sets the signal source of a PWM waveform.
PM, DEVI
<pm phase offset>
Sets the phase offset of a PM waveform. Only applicable when the signal source is set to INT.
ASK, SRC
<src>
Sets the signal source of an ASK waveform.
ASK, KFRQ
<ask key frequency>
Sets the key frequency of an ASK waveform. Only applicable when the signal source is set to
INT.
INT.
FSK, KFRQ
<fsk frequency>
Sets the frequency of an FSK waveform. Only applicable when the signal source is set to INT.
FSK, HFRQ
<fsk hop frequency>
Sets the jump frequency of an FSK waveform.
FSK, SRC
<src>
Sets the signal source of an FSK waveform.
CARR, WVTP
<wave type>
Sets the carrier waveform type value.
CARR, FRQ
<frequency>
Sets the carrier waveform frequency value.
CARR, AMP
<amplifier>
Sets the carrier waveform amplification value.
CARR, OFST
<offset>
Sets the carrier waveform offset value.
CARR, SYM
<symmetry>
Sets the carrier waveform symmetry value.
CARR, DUTY
<duty>
Only applying to Square carrier waveforms, this parameter sets the duty cycle value.
CARR, PHSE
<phase>
Sets the carrier waveform phase value.
PLEASE NOTE THE FOLLOWING:
The Mod function is disabled when Pulse or Noise waveform types are selected.