Manual Mode 
 
In this mode, a pair of cursors will appear. You can adjust the cursors manually to 
measure the X (or Y), X increment (or Y increment) between cursors and the 
reciprocal of X increment on the waveform of the selected source (CH1, CH2 or 
MATH). 
 
Press Cousor  Mode  “Manual” to turn the manual cursor function on and the 
measurement results will be displayed at the upper left corner of the screen in the 
following mode. 
 
 
  The X or Y value at cursor A (AX or AY): X value takes the trigger position as 
reference and Y value takes the channel GND as reference.   
  The X or Y value at cursor B (BX or BY): X value takes the trigger position as 
reference and Y value takes the channel GND as reference.   
  The horizontal or vertical difference between cursor A and B (BX-AX or BY-AY).   
  The reciprocal of the horizontal difference between cursor A and B (1/|dX|). 
 
If needed, please refer to the following steps to modify the parameters of manual 
cursor measurement. 
 
1.  Select Cursor Type 
Press Type to select “X” or “Y”. The X cursors are a vertical dotted line (cursor A) 
and a vertical solid line (cursor B) and are usually used to measure time 
parameters. The voltage cursors are a horizontal dotted line (cursor A) and a 
horizontal solid line (cursor B) and are usually used to measure voltage 
parameters. 
   
2.  Select Measurement Source 
Press Source to select the waveform of the analog channels (CH1-CH4) or math 
operation results (MATH) for measurement.