Rigol DS1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, DS1074Z-S Data Sheet

Product codes
DS1074Z-S
Page of 264
RIGOL 
DS1000Z User’s Guide 
12-12 
To Output Arbitrary Waveform 
 
DS2000A allow users to define arbitrary waveforms and to save them to the internal 
or external memory. The internal memory can store 10 arbitrary waveforms at most. 
1 to 16384 points can be included in the user-defined waveforms, namely 1 pts to 16 
kpts。Press Src1 Setup to enter the waveform setting interface. Press Wave to 
select “Arb”, at this point, you can edit arbitrary waveforms. 
 
1.  Output   
Press Output to turn the signal output on or off. 
 
2.  To set frequency 
Press Frequency to set the frequency of the current signal and the range 
available is from 100 mHz to 1 MHz.   
Method one: 
Rotate 
  directly to set the desired frequency value. 
Method two: 
Press 
  to input desired paramrter values and unist directly using the pop-up 
numeric keyboard. 
 
3.  To set amplitude   
Press Amplitude to set the amplitude of the current signal and the range 
available is from 20 mVpp to 5Vpp (when the impedance is “HighZ”) or from 10 
mVpp to 2.5 Vpp (when the impedance is “50 Ω”). 
Method one: 
Rotate 
  directly to set the desired implitude value. 
Method two: 
Press 
  to input desired paramrter values and unist directly using the pop-up 
numeric keyboard. 
 
4.  To set the DC offset voltage 
Press Offset to set the DC offset of the current signal and the range available is 
from “-2.5 V + the amplitude value currently set/2” (when the impedance is 
“HighZ”) to “2.5 V + the amplitude value currently set/2” (when the impedance is 
“50 Ω”). 
Rotate 
  directly to set the desired DC offset value.