Probe Compensation 
 
When the probes are used for the first time, you should compensate the probes to 
match the input channels of the oscilloscope. Non-compensated or poorly 
compensated probes may cause measurement inaccuracy or error. The probe 
compensation procedures are as follows. 
 
1.  Perform steps 1, 2, 3 and 4 of 
 in the previous section.   
2.  Check the waveforms displayed and compare them with the following. 
         
 
 
 
 
Figure 0-8 Probe Compensation 
 
3.  Use a nonmetallic driver to adjust the low-frequency compensation adjustment 
hole on the probe until the waveform displayed is as the “Perfectly 
compensated” in the figure above.   
 
 
 
 
Over compensated        Perfectly compensated     Under compensated