Channel Coupling 
 
Set the coupling mode to filter out the undesired signals. For example, the signal 
under test is a square waveform with DC offset. 
  When the coupling mode is “DC”: the DC and AC components of the signal under 
test can both pass the channel. 
  When the coupling mode is “AC”: the DC components of the signal under test 
are blocked. 
  When the coupling mode is “GND”: the DC and AC components of the signal 
under test are both blocked. 
 
Press CH1  Coupling and use 
  to select the desired coupling mode (the 
default is DC). The current coupling mode is displayed in the channel label at the 
bottom of the screen. You can also press Coupling continuously to switch the 
coupling mode.   
 
 
Bandwidth Limit 
 
Set the bandwidth limit to reduce display noise. For example, the signal under test is 
a pulse with high frequency oscillation. 
  When bandwidth limit is disabled, the high frequency components of the signal 
under test can pass the channel. 
  Enable bandwidth limit and limit the bandwidth to 20 MHz, the high frequency 
components that exceed 20 MHz are attenuated. 
 
Press CH1  BW Limit and use 
  to enable or disable bandwidth limit (the 
default is OFF). When bandwidth limit (20 MHz) is enabled, the character “B” will be 
displayed in the channel label at the bottom of the screen. You can also press BW 
Limit continuously to switch between on and off of the bandwidth limit.