DS1000Z User’s Guide
 
 
3-5  
 
XY Mode 
 
In this mode, the oscilloscope changes the two channels from voltage-time display 
mode to voltage-voltage display mode. The phase deviation between two signals 
with the same frequency can be easily measured via Lissajous method. The figure 
below shows the measurement schematic diagram of the phase deviation. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
According to sinθ=A/B or C/D (wherein, θ is the phase deviation angle between 
the two channels and the definitions of A, B, C and D are as shown in the figure 
above), the phase deviation angle is obtained, that is: 
θ=±arcsin (A/B) or ±arcsin (C/D) 
If the principal axis of the ellipse is within quadrant I and III, the phase deviation 
angle obtained should be within quadrant I and IV, namely within (0 to π/2) or (3π/2 
to 2π). If the principal axis of the ellipse is within quadrant II and IV, the phase 
deviation angle obtained should be within quadrant II and III, namely within (π/2 to 
π) or (π to 3π/2).   
 
XY function can be used to measure the phase deviation occurred when the signal 
under test passes through a circuit network. Connect the oscilloscope to the circuit to 
monitor the input and output signals of the circuit. 
II  I 
III  IV 
A  B 
The signal must be 
centered horizontally      
 
II  I 
III  IV