Rigol DS1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, DS1074Z-S Data Sheet

Product codes
DS1074Z-S
Page of 264
RIGOL 
DS1000Z User’s Guide 
5-6 
Trigger Holdoff 
 
Trigger holdoff can be used to stably trigger the complex waveforms (such as pulse 
series). Holdoff time is the amount of time that the oscilloscope waits before 
re-arming the trigger circuitry. The oscilloscope will not trigger until the holdoff time 
expires.   
 
 
 
Press MENU  Setting  Holdoff in the trigger control area (TRIGGER) at the 
front panel and use 
  to modify the holdoff time (the default is 16 ns) until the 
waveform triggers stably. The adjustable range of holdoff time is from 16 ns to 10 s. 
Note that trigger holdoff is not available for video trigger. 
 
 
 
 
 
 
 
 
 
 
 
 
 
释抑时间 
 
触发位置