To determine the hottest spot on the measured object you have to “scan” the object with measuring button (7) 
held down, performing systematic zigzag movements until you find the hottest spot. The highest measured 
temperature is shown in the upper right of the display as maximum temperature (if Max function is selected).
In  order  to  obtain  exact  measured  values,  the  infrared  thermometer  has  to  be  adjusted  to  the  ambient 
temperature. Allow the product to adjust to the ambient temperature after relocation.
Shiny surfaces affect the measured result. To compensate, the shiny part of the surface can be covered with 
adhesive tape or matt black paint. The product cannot take measurements through transparent surfaces such 
as glass. Instead, it measures the surface temperature of the glass.
Longer measurements of high temperatures, with close measuring distance, lead to self-heating of the 
measuring  device  and  thus  to  inaccurate  measurements.  In  order  to  obtain  exact  measured  values, 
remember the following rule of thumb: The higher the temperature the greater the measuring distance and 
the shorter the measuring time.
c)  IR measuring spot size - Distance to surface level (Distance to spot level; D/S)
In order to obtain precise measured results, the object to be measured must be larger than the measuring IR spot of 
the IR thermometer. The temperature recorded is the average temperature of the area measured. The smaller the 
object to be measured, the shorter the distance must be to the infrared thermometer. The exact size of the measuring 
spot is indicated in the following diagram. It is also printed on the device. For precise measurements, the object to be 
measured should be at least twice the size of the measuring spot.
d) Target laser
The target laser can be activated or deactivated. Turn the measuring device on by pressing the measuring button (7). 
Press the “Laser/background illumination” button (3). When the laser is activated, the laser display (triangle with 
radiation symbol) is shown in the display. To deactivate the laser, press the “Laser/background illumination” button (3) 
until the laser display goes out.
e) Background illumination
With the measuring device running, use the button “Laser/Background illumination” (3) to activate/deactivate the 
display’s background illumination. Press this button, until the background illumination is activated or deactivated.
SETTINGS mENU
The product is equipped with a setting menu. To get into the menu, press and hold down the “MODE” button (6) while 
the product is turned on until the degree of emission flashes in the display. Now press the “MODE” button, until you 
get to the desired settings function in the menu. The values on the settings menu can be changed by using the “up” 
button (2) and the “down” button (8). Confirm your input with the measuring button (7), or press the “MODE” button (6) 
in order to move to the next setting. When the device is switched off, the set values remain stored.
a) min/max measurement
The product can indicate in the secondary display (15) the lowest (MIN) or the highest (MAX) temperature value that 
occurs during the measurement. In the setting menu, press the “MODE” button (6) until the display “MIN” (19) or 
“MAX” (13) starts flashing in the display. Select the desired display type for the secondary display (15) (MIN or MAX), 
by using the buttons “up” (2) and “down” (8).
b) “°c/°f” conversion between measurement units
Keep pressing the “MODE” button (6) in the setting menu, until the measuring unit (20) begins to flash. With the “up” 
button (2) and the “down” button (8) the measuring unit can be changed from °C (degrees Celsius) to °F (degrees 
Fahrenheit).
c) Setting the emission level
The measuring device allows the emission level to be set. Thus, different materials and surfaces can be measured 
precisely (also refer to chapter “a) Operating principle”).
Keep pressing the “MODE” button (6), until the symbol for the emission level (14) starts flashing in the display. Using 
the buttons “up” (2) and “down” (8) the emission level can be adjusted from 0.10 to 1.00 depending on the object 
to be measured.
Following the technical data you will find a table with typical materials and their emission level. Many 
organic materials have an emission level of 0.95. Thus the default setting for the emission level is 0.95. 
The emission level of a surface can be measured with a conventional thermometer with a special surface 
sensor. Measure the temperature of the surface to be measured with a conventional thermometer. 
Change the emission level of the infrared thermometer until the measured value (18) matches that of the 
conventional surface measurement. This, however, is only necessary for highly precise measurements.
d) Alarm mode
The measuring device has an alarm function for cases where a set temperature value is exceeded. The alarm is 
indicated by a beeper and the background illumination flashing red. Due to this function, the measuring device is ideal 
for monitoring, etc. The alarm is triggered when the set alarm value is exceeded.
Keep pressing the “MODE” button (6), until the symbol for alarm “SET HIGH” (17) flashes and a temperature value 
is shown in the display. Using the buttons “up” (2) and “down” (8) set the alarm value that will trigger the alarm if it 
is exceeded.
The acoustic alarm (beeper) can be deactivated. Keep pressing the “MODE” button (6) in the settings menu, until the 
symbol for the visual alarm (16) starts flashing in the display. The visual alarm can be activated or deactivated using 
the buttons “up” (2) and “down” (8).
mAINTENANcE AND cARE
a) cleaning of the lens
Remove loose particles with clean compressed air and wipe off remaining residues with a fine brush. Clean the 
surface with a lens cleaning cloth or a clean, soft, lint-free cloth. To remove fingerprints and other fat residues, the 
cloth can be moistened with water or lens cleaning solution. Do not use any acidic, alcoholic or other solvents or 
rough, linty cloth to clean the lens. Avoid applying too much pressure during cleaning.
b) cleaning the housing
Use a moist cloth to clean the enclosure. Dry the product after cleaning. Do not use abrasive agents or solvents!
c) Protective bag
Store the thermometer after use and cleaning in the protective bag. The protective bag protects against dirt, dust 
and damage.
DISPOSAL
a) Product
Electronic devices are recyclable waste and must not be disposed of in the household waste.
At the end of its service life, dispose of the product according to the relevant statutory regulations.
Remove any inserted (rechargeable) batteries and dispose of them separately from the product.
b) (Rechargeable) batteries
You as the end user are required by law (Battery Ordinance) to return all used batteries/rechargeable batteries. 
Disposing of them in the household waste is prohibited.
Contaminated  (rechargeable)  batteries  are  labelled  with  this  symbol  to  indicate  that  disposal  in  the 
domestic waste is forbidden. The designations for the heavy metals involved are: Cd = Cadmium, Hg = 
Mercury, Pb = Lead (name on (rechargeable) batteries, e.g. below the trash icon on the left).
Used (rechargeable) batteries can be returned to collection points in your municipality, our stores or 
wherever (rechargeable) batteries are sold.
You thus fulfil your statutory obligations and contribute to the protection of the environment.
TEchNIcAL DATA
Operating Voltage:  .............................................9 V/DC (9 V block battery)
Response time:  ..................................................<1 second
Spectrum:  ..........................................................6 – 14 μm
Emission level:  ...................................................0.1 to 1.00 adjustable
Definition  ............................................................0.1 °C
IR-measuring spot size:  .....................................IR 10:1
Laser:  ................................................................. Rated power <1 mW, laser class 2, Wave length 630 – 670 nm
Operating temperature:  ..................................... 0 to +50 °C
Operating humidity:  ..............................................10 to 90 % RH
Storage temperature:  .........................................-10 to +60 °C
Storage humidity:  ...............................................10 to 80 % RH
Weight:  ..............................................................180 g
Dimensions (W x H x D):  ...................................42 x 160 x 82 mm
Accuracy (for ambient temperatures of 23 – 25 °C, emission level 0.95):
Temperature measuring range
Accuracy
-50 to 0 °C
± 5 °C
0 to +500 °C
±3% of the measurement value or ±3 °C
a) Emission level of different surfaces
The emission levels in the table are approximate values. Different parameters such as geometry and the 
surface quality can affect the emission level of an object.
Surface
Emission level
Surface
Emission level
Asphalt
0.90 – 0.98
Varnish (matt)
0.97
Concrete
0.94
Human skin
0.98
Ice
0.96 – 0.98
Mortar
0.89 – 0.91
Ferric oxide
0.78 – 0.82
Paper
0.70 – 0.94
Soil. Humus
0.92 – 0.96
Plastics
0.85 – 0.95
Hard plaster
0.80 – 0.90
Sand
0.90
Glass/ceramics
0.90 – 0.95
Textiles
0.90
Rubber (black)
0.94
Water
0.92 – 0.96
Varnish
0.80 – 0.95
Bricks
0.93 – 0.96