Greisinger EASYLOG 80 CL Humidity: 0 - 100 % r.H · Temperature: -25 to +60 °C · Pressure: 300 - 1100 hPa Temperature Dat 603414 Data Sheet

Product codes
603414
Page of 12
E40.0.01.6C-02 Operating 
Manual 
EASY
LOG
 80 CL 
Page 12 of 12 
12.  Frequently asked questions [FAQ] 
Problem: 
The logger cannot be stopped via keys.
 
Possible reason: 
“Logger stop via buttons disabled” has been activated at the configu-
ration of the logger. 
Remedy: 
Deactivate this function by means of GSOFT 40K (at tab „settings“).  
Problem: 
The logger cannot be stopped both via keys and GSOFT 40K. 
Possible reason: 
“Do not stop active logger” has been activated at the configuration of 
the logger. 
Remedy: 
Deactivate this function by means of EASYBus-configurator or 
EbxKonfig. 
Problem: 
The alarm settings or the sea-level correction cannot be 
changed. 
Possible reason: 
The logger has still saved data. 
Remedy: 
Stop logger as recommended by the software (delete data).
 
Problem: 
The desired calculated variable cannot be changed. 
Possible reason: 
The logger has still saved data. 
Remedy: 
Stop logger as recommended by the software (delete data).
 
Problem: 
A error message is displayed. 
Error message 
Possible reason 
procedure / remedy 
 
measuring range 
exceeded 
Measuring range of the chan-
nel has been exceeded. 
Operate logger within its specifica-
tions.  
humidity: sensor bedewed 
Twist filter cap carefully off and let 
sensor dry.   
Sensor defective 
Send the logger to the manufac-
turer for repair. 
 
measuring range 
under-run
 
Measuring range of this chan-
nel is under-run.
 
Operate logger within its specifica-
tions.
 
Sensor defective
 
Send the logger to the manufac-
turer for repair.
 
 
system error
 
Recording error 
Reset the error via GSOFT 40K.
 
If the error message is displayed 
furthermore send the logger to the 
manufacturer for repair.
 
System error 
Send the logger to the manufac-
turer for repair.
 
 
sensor error
 
Humidity / temperature sensor 
has moved out of its socket  
Check the position of humidity / 
temperature sensor. 
Sensor defective
 
Send the logger to the manufac-
turer for repair.
 
 
calculation error
 
Error message at a variable 
needed for the calculation 
(humidity / temperature / pres-
sure) 
Remedy the reason for error at this 
channel.