Chauvin Arnoux P01157050 Simple Data Logger P01157050 User Manual

Product codes
P01157050
Page of 172
SIMPLE LOGGER II 
 
 
 
English 
 
 
Page 39 
2. PRESENTATION 
2.1  Description 
 
The Simple Logger
®
  II is a self-contained digital recorder including one or two analogue input acquisition 
channels depending on the model (powered from a set of two alkaline batteries). 
The analogue input signal is sampled and converted into digital signals. The data obtained are processed, 
time-stamped, and recorded. An optically isolated USB (Universal Serial Bus) port allows the downloading of 
the recorded data to a desktop computer. 
The main advantage of this recorder is that it makes it possible to solve many recording problems using the 
simple and intuitive DataView
®
 software interface. 
For the measurement of AC signals, the frequency of the signal is followed (within of ± 2Hz of the 50 or 60Hz 
nominal network frequency) and 64 samples measured for each alternation. 
For a TRMS* measurement, the computation is done on the basis of the 64 samples of one alternation; this 
process is repeated 8 times a second. Information about harmonics is calculated by DataView
®
  software 
from the data of these 64 samples (so the input signals during the intervals separating two measurements 
are not taken into account). 
DC signal measurements are made 8 times per recording interval. 
*: TRMS = True Root Mean Square (true RMS value). 
 
2.2  Composition 
2.2.1  Standard models