Chauvin Arnoux P01157060 Simple Data Logger P01157060 User Manual
Product codes
P01157060
SIMPLE LOGGER II
English
Page 51
MODEL
L481
Electrical
Channels
One
Input
Two recessed 4mm safety banana jacks
Input Level
-850VDC to +850VDC
Accuracy
(50/60 Hz)
(50/60 Hz)
0 to 5 V: unspecified
5 to 50 V: ±(0.5% of Reading +1 V)
50 to 850 V: ±(0.5% of Reading +0.5 V)
Resolution
0.1 V
Maximum Input Voltage***
±1,020 V
DC
Input Impedance
>150 k
Ω
Sample Rate
Maximum of 8 per second
Storage Rate
Programmable from 12 ms to 1 per day
Storage Modes
Start/Stop, FIFO and Extended Recording Mode* (XRM™)
Recording Length
15 minutes to 8 weeks, programmable using DataView
®
Memory
240,000 measurements (512 KB). Recorded data is stored in non-volatile
memory and retained even if battery is low or removed.
Communication
USB 2.0 optically isolated
Power Source**
2 x 1.5 V AA (LR6) alkaline batteries
Battery Life
100 hours to > 45 days
(dependent on storage rate/recording length)
Mechanical
Dimensions
4.94 x 2.75 x 1.28" (125 x 70 x 32 mm)
Weight (with battery)
6.4 oz (180 g)
Case
UL94-V0
Vibration
IEC 68-2-6 (1.5 mm, 10 to 55 Hz)
Shock
IEC 68-2-27 (30 G)
Drop
IEC 68-2-32 (1 m)
Environmental
Operating Temperature
14° to 122°F (-10° to 50°C)
Storage Temperature
-4° to 140°F (-20° to 60°C)
Relative Humidity
Up to 85% at 95°F (35°C), Non-condensing
Altitude
2000 m
Safety & Electro-magnetic compatibility
Safety Rating
EN61010-1; 600 V CAT III, 300 V CAT IV; Pollution Degree 2
Protection Degree
IP40
CE Approved
Yes
*See the end of chapter