Chauvin Arnoux P01157060 Simple Data Logger P01157060 User Manual

Product codes
P01157060
Page of 172
SIMPLE LOGGER II 
 
 
 
English 
 
 
Page 51 
 
MODEL 
L481 
Electrical 
Channels 
One 
Input 
Two recessed 4mm safety banana jacks 
Input Level 
-850VDC to +850VDC 
Accuracy 
(50/60 Hz) 
0 to 5 V: unspecified 
5 to 50 V: ±(0.5% of Reading +1 V) 
50 to 850 V: ±(0.5% of Reading +0.5 V) 
Resolution 
0.1 V 
Maximum Input Voltage*** 
±1,020 V
DC
 
Input Impedance 
>150 k
Ω 
Sample Rate 
Maximum of 8 per second 
Storage Rate 
Programmable from 12 ms to 1 per day 
Storage Modes 
Start/Stop, FIFO and Extended Recording Mode* (XRM™) 
Recording Length 
15 minutes to 8 weeks, programmable using DataView
®
 
Memory 
240,000 measurements (512 KB). Recorded data is stored in non-volatile 
memory and retained even if battery is low or removed. 
Communication 
USB 2.0 optically isolated 
Power Source** 
2 x 1.5 V AA (LR6) alkaline batteries 
Battery Life 
100 hours to > 45 days 
(dependent on storage rate/recording length) 
Mechanical 
Dimensions 
4.94 x 2.75 x 1.28" (125 x 70 x 32 mm) 
Weight (with battery) 
6.4 oz (180 g) 
Case 
UL94-V0 
Vibration 
IEC 68-2-6 (1.5 mm, 10 to 55 Hz) 
Shock 
IEC 68-2-27 (30 G) 
Drop 
IEC 68-2-32 (1 m) 
Environmental 
Operating Temperature 
14° to 122°F (-10° to 50°C) 
Storage Temperature 
-4° to 140°F (-20° to 60°C) 
Relative Humidity 
Up to 85% at 95°F (35°C), Non-condensing 
Altitude 
2000 m 
Safety & Electro-magnetic compatibility 
Safety Rating 
EN61010-1; 600 V CAT III, 300 V CAT IV; Pollution Degree 2 
Protection Degree 
IP40 
CE Approved 
Yes 
*See the end of chapter