3. Yellow LED
ORDER
Lit
CLEARS THE ALARM DISPLAY
(see § 4.6)
STATUS
OFF
No alarm detected (all channels)
Single flash
There has been an alarm in at least one
channel at least once
channel at least once
Double flash
There is currently an alarm in at least one
channel
channel
Rapid flash
Ready to clear alarm indications
4. Red LED
ORDER
Lit
ERASES THE MEMORY
(see § 4.7)
STATUS
OFF
No data in memory
Single flash
Memory partially filled
Double flash
Memory full
Rapid flash
Ready to erase the memory
5. Blue LED
ORDER
Lit
STOPS THE DEVICE
STATUS
OFF
Battery voltage OK (> 2.2 volts)
Single flash
Batteries worn, must be replaced (< 2.2
volts)
volts)
Double flash
Confirms the programming of a recording
* To determine whether the device is off or in standby mode, press the "PRESS" button for 0.5 second: if all
of the LEDs light, the device is not off.
Note: The overload warning appears as soon as an input quantity exceeds by 10% the
specified value of the input range.
When the battery voltage falls below 1.7 volt, the device stops automatically (if a recording is in progress, it is
stopped and the data are saved).
stopped and the data are saved).
Sleep mode: The device switches to low-consumption mode after one minute with no press of the button
and remains in this mode for as long as the button is not pressed or until the time at which programmed
recording is to be started is reached.
Recording mode: The device switches to low-consumption mode between measurement acquisition stages.
The lower the recording frequency, the larger the fraction of the time spent in low-consumption mode, and so
the longer the possible total recording time.
OFF mode: The device is in low-consumption mode; it will switch on automatically to start programmed
recording.