Chauvin Arnoux P01157040 Voltage Simple Data Logger P01157040 User Manual

Product codes
P01157040
Page of 172
SIMPLE LOGGER II 
 
 
 
English 
 
 
Page 42 
6.  "Flash Upgrade" switches (not shown) 
These two switches, placed near the "RESET" button, are used to take control of the device if an upgrade of 
the internal operating program fails. 
 
 
WARNING:  if the "RESET" button is actuated while recording is in progress, recording will be 
interrupted and the data in memory may be lost. 
 
 
 
 
3. CHARACTERISTICS 
 
Reference conditions: 23°C ± 3°C, 30-50% RH, DC or 50/60Hz, no external AC magnetic field, DC magnetic 
field 
≤ 40 A / m, conductor centred, supply voltage: 3 V ± 10%. 
 
MODEL 
L101 
L102 
Electrical 
Inputs 
One 
Two 
Input connection 
BNC 
Input level 
0 to 1 V
AC
 
Uncertainties(50 / 60 Hz) 
 
 
Resolution 
0 to 10 mV: unspecified 
10 to 50 mV: ± (0.5% of reading + 1 mV) 
50 to 1,000 mV: ± ( 0,5 mV + 0,5 mV) 
0,1 mV
AC
 
Maximum input level 
5 Vrms or ± 7.07 Vpeak 
Input impedance 
800 k
Ω 
Sample rate 
64 samples/cycle 
Storage rate 
Programmable from 125 ms to 1 day 
Storage modes 
Start / Stop, FIFO or Extended Recording Mode* (XRM™) 
Recording length 
Programmable using DataView
®
 from 15 minutes to 8 weeks 
Memory 
240,000 measurements (512 kB). The recorded data is stored in non-volatile 
memory retained even if the battery is low or removed. 
Communication 
USB 2.0 optically isolated 
Power source ** 
2 x 1.5 V AA (LR6) cell Alkaline batteries 
Battery life 
100 hours to >45 days (dependent on storage rate/recording length) 
Mechanical 
Dimensions 
136x70x32 mm (5.38x2.75x1.28") 
Weight (with batteries) 
180 g (6.4 oz) 
Case 
UL94-V0 
Vibration 
IEC 68-2-6 (1.5 mm, 10 to 55 Hz) 
Shock 
IEC 68-2-27 (30 G) 
Drop 
IEC 68-2-32 (1 m) 
Environmental 
Operating temperature 
- 10°C to + 50°C (14°F to 122°F) 
Storage temperature 
- 20°C to + 60°C (- 4°F to + 140°F) 
Relative humidity 
up to 85% at 35°C (95°F) without condensation 
Altitude 
2,000 m 
Safety 
Conformity 
EN 61010-1; 50 V CAT III; Pollution level 2 
Level of protection 
IP 40 
*See the end of chapter