Chauvin Arnoux P01157040 Voltage Simple Data Logger P01157040 User Manual

Product codes
P01157040
Page of 172
SIMPLE LOGGER II 
 
 
 
English 
 
 
Page 64 
5. MAINTENANCE 
 
All of the instruments undergo mechanical and electronic checks before shipping. Every precaution is taken 
to ensure that you receive an undamaged instrument. If there is any damage, inform the carrier immediately. 
The manufacturer cannot be held liable for any incident, accident, or malfunction following a repair done 
other than by its customer service department or an approved repairer. 
 
5.1  Replacement of the batteries 
 
 
 
 
WARNING: Risk of electric shock. Disconnect all inputs or remove the clamp from 
any conductor before opening the battery compartment cover. To eliminate the risk 
of loss of recorded data, switch the device off before changing the batteries. 
 
 
Disconnect all inputs or remove the clamp from any conductor before opening. 
Switch the device off 
Remove the screw holding the cover, press on the cover and slide it off. 
Replace the two 1.5 V AA (LR6) alkaline batteries. Do not remove both batteries at once; replace only the 
first one completely, then the other. (The maintenance capacitor can maintain a supply for approximately 10 
seconds while the batteries are being replaced.) 
Replace the battery compartment cover and the screw 
Keep the "PRESS" button pressed for 2 seconds to restart the device. 
 
 
NOTE:  If the device is left too long without batteries, it will be necessary, after replacing the 
batteries, to reset the date and time from the control panel of the Simple Logger
®
 II. 
Leave the batteries in the device only for short periods of time. If the device is to be left unused for an 
extended period, we recommend removing the batteries. 
When replacing the batteries, always replace them both. Never leave an old battery in place next to a new 
one. 
 
 
5.2  Replacing the fuse (model L111) 
 
 
 
 
WARNING:  Risk of electric shock. Remove the clamp from any conductor and 
disconnect all inputs before replacing the fuse. 
 
 
Using a flat blade screwdriver, press the cap of the fuse holder and unscrew a quarter turn; 
When you stop pressing, the cap and fuse pop out by themselves; 
Replace the fuse with an identical model: FA, 2 A / 250 V (5 x 20 mm); 
Replace the cap and screw it back in a quarter turn.