Gw Instek GPT-9601 Insulation measuring device, GPT-9601 User Manual
Product codes
GPT-9601
GPT-9600 Series User Manual
52
PASS / FAIL Test Judgments for Auto Mode
Background
The test results for the Auto Mode testing are
identical as the results from a single test,
however:
identical as the results from a single test,
however:
If one of the tests fails, then the FAIL
judgment for that test will be shown on the
display, even if the other test is judged as
PASS.
judgment for that test will be shown on the
display, even if the other test is judged as
PASS.
Only one test result can be viewed at any one
time.
time.
Note
A test will be judged PASS when:
The H and L limits (high and low limits)
have not been tripped during the test time.
have not been tripped during the test time.
A test will be judged FAIL when:
Either the H or L limit has been tripped
during the test time.
during the test time.
The ARC current limit has been tripped
during a test (if enabled).
during a test (if enabled).
View Test
Results~PASS
Results~PASS
1. When both tests are judged as
PASS, PASS will be displayed for
the last test, the buzzer will sound
once and the PASS indicator will
be lit green.
the last test, the buzzer will sound
once and the PASS indicator will
be lit green.
PASS
H = 2 2 . 0 m A
C
m A
L
0
=
0 0
.
6 0 S
= 0
T
CW
D
A CW
R
I
-
R A
I
T M E
I
0 90
k V
S
A
P
S
A
m
1 00
2. Turn the scroll wheel left to view
the first test result. Turn the scroll
wheel back to the right to view the
last test result again.