EXTERNAL CONTROL 
61 
Remote Controller Operation 
 
Description 
The GPT-9600 accepts external remote 
controllers with a START and STOP button. To 
use the REMOTE terminal, the  
GPT-9600 must first be configured to accept a 
remote controller.  
Operating a remote controller is the same as 
operating the START and STOP buttons on the 
front panel. 
Steps 
1.  Insert the lead of remote controller 
into the REMOTE terminal. 
REMOTE
  
 
2.  Configure the Start Ctrl option to 
REMOTE CONNECT in the 
Common Utility menu. 
Page 56 
 
3.  The tester will now only be able to 
start a test using a remote 
controller. 
 
 NOTE 
Even if the GPT-9600 is configured to use the 
REMOTE CONNECT option, the STOP button on 
the front panel can still be used to stop a test.   
 
4.  To return the operation control to 
the front panel, configure the Start 
Ctrl option to FRONT PANEL. 