Gw Instek GPT-9802 Insulation measuring device, GPT-9802 User Manual
Product codes
GPT-9802
GPT-9000 Series User Manual
48
3. Use the scroll wheel to set the ARC
mode.
ARC MODES:
OFF, ON AND CONTINUE,
ON AND STOP
ON AND STOP
4. Press the EDIT/SAVE key to save
and exit the MANU Utility and go
back to EDIT status.
back to EDIT status.
EDIT/SAVE
Note
The ESC key can be pressed at any time in the
Utility menu to cancel and exit.
Utility menu to cancel and exit.
5. If the ARC MODE was set to either ON AND
CONTINUE, or ON AND STOP, the ARC
current level can be edited.
current level can be edited.
6. Use the UP/DOWN arrow keys to
move the cursor to the ARC setting.
7. Use the scroll wheel to edit the
ARC level.
ACW
DCW
DCW
1.000mA~080.0mA
1.000mA~020.0mA
1.000mA~020.0mA
Note
The ARC setting range is directly related to the HI
SET current limit.
SET current limit.
ACW
HI SET Limit
ARC range
0.001mA~0.999mA
1.000mA ~2.000mA
01.00mA~09.99mA
01.00mA ~20.00mA
010.0mA~042.0mA
001.0mA ~080.0mA