Gw Instek GPT-9904 Insulation measuring device, GPT-9904 User Manual

Product codes
GPT-9904
Page of 171
 
 
GPT-9000 Series User Manual 
122
 
SYSTem:BUZZer:FTIMe 
Set
Query
 
 
Description 
Sets the FAIL Sound duration in seconds.  
Syntax 
Query Syntax 
SYSTem:BUZZer:FTIMe <NR2> 
SYSTem:BUZZer:FTIMe? 
Parameter/ 
Return parameter 
<NR2> 
0.2~999.9 
Example 
SYST:BUZZ:FTIM 1 
Sets the buzzer to 1 second for a FAIL judgment. 
SYSTem:ERRor 
Query
 
 
Description 
Returns any errors in the output buffer. See the 
error code table below for details. 
Query Syntax 
SYSTem:ERRor ? 
Return parameter  <string> 
Returns an error string that includes 
an error code and an error 
description. 
 
 
Error Code Table 
 
Error code, Error description 
 
0,No Error 
 
20,Command Error 
 
 
21,Volume Error 
 
 
22,String Error 
 
 
23,Query Error 
 
 
24,Mode Error 
 
 
25,Time Error 
 
 
26,DC Over 50W 
 
 
27,GBV > 5.4V 
 
 
30,Voltage Setting Error 
 
 
31,Current Setting Error 
 
 
32,Current HI SET Error 
 
 
33,Current LOW SET Error 
 
 
34,Resistance HI SET Error 
 
 
35,Resistance HI SET Error